The 8 Series Sampling Oscilloscope provides a comprehensive optical test solution for telecom and datacom applications, as well as general purpose optical component testing. The TSO8C17 and TSO8C18 optical modules provide > 30 GHz optical bandwidth, plus fully integrated Optical Reference Receivers (ORR)1 enabling both single mode and multi-mode conformance testing at 850 nm, 1310 nm, and 1550 nm bands. Key performance specifications • Optical bandwidth above 30 GHz • Single mode and multi-mode support for short and long wavelength optical testing • Optical Reference Receiver (ORR)1 support for standard compliance testing Applications • Design/verification of High-Speed Components and Systems • Signal integrity analysis • Compliance test for industry NRZ and PAM4 standards: 10G, 50G, 100G, 200G, 400G IEEE 802.3TM standards (such as 400GBASEFR8, 400GBASEDR4) and similar optical direct detect standards.
Key features • New system architecture • Disaggregated: The product consists of the TSO820 mainframe, pluggable modules, acquisition hardware, and the TSOVu software analysis application that runs on a user's PC with Windows. Users have the capability to scale their analysis platforms to their needs and can connect from anywhere on the network • Configurable: The TSO820 mainframe supports userswappable current and future optical modules • Optical modules • Accurate testing and characterization of short or long wave optical signals using the high sensitivity and low noise performance of the TSO8C17 or TSO8C18 modules • Optical reference receivers (ORR)1 supports specified requirements for standards-mandated compliance testing • User configurable bandwidth supported through Bandwidth Enhancement (BWE) filters. Extend the TSO8C17 and TSO8C18 optical bandwidth up to 35 GHz in 0.0001 GHz increments for precise control • Extinction ratio measurements with built-in variable ER correction to ensure accuracy and repeatability • Analysis with TSOVu® • TSOVu: Oscilloscope software application that runs independent of the oscilloscope mainframe on users' computers or server for both live and post-processing of acquired data • TSOVu offers comprehensive analysis of NRZ or PAM4 optical signals. Includes support for eye diagrams, NRZ mask testing, optical measurements such as TDECQ, and other standard measurements • New measurement plug-ins can be dynamically installed as needed • High test throughput • Simultaneous capture of all channels at a high sample acquisition rate of 300 kS/s • Sophisticated Programmatic Interface (PI) for automation environments enables the highest test throughput. Each
Bandwidth4 | 850 nm | 1310 nm | 1550 nm |
---|---|---|---|
12.6 GHz | 4.2 μW | 2.8 μW | 3.0 μW |
13.28125 GHz | 4.3 μW | 2.9 μW | 3.0 μW |
19.335 GHz | 5.3 μW | 3.7 μW | 3.9 μW |
21 GHz | 6.2 μW | 4.2 μW | 4.4 μW |
22.5 GHz | 8.1 μW | 5.0 μW | 5.4 μW |
Bandwidth4 | 850 nm | 1310 nm | 1550 nm |
---|---|---|---|
12.6 GHz | 6.0 μW | 3.6 μW | 3.9 μW |
13.28125 GHz | 6.0 μW | 3.6 μW | 3.9 μW |
19.335 GHz | 7.5 μW | 4.5 μW | 4.8 μW |
21 GHz | 8.3 μW | 5.0 μW | 5.4 μW |
22.5 GHz | 11.1 μW | 6.7 μW | 6.9 μW |